|
STERILE
PROCESSING UNIVERSITY
"Alternatives to Steam Sterilization" Copyright © 2011 - Sterile Processing University, LLC - All Rights Reserved. This in-service (including photos) may not be copied or used without permission of the author. Updated 9.21.11 **This
site is best viewed at 1024x768 resolution on your monitor. Ethylene oxide (EO) gas is a chemical agent used for the sterilization of heat, pressure and/or moisture sensitive items. It is a member of the ether family also known as peroxide. EO is a liquid that, at room temperature, becomes a gas. It is used as a fumigant, pesticide and for sterilization.
There
are several types of EO used for sterilization. The most common is the
100% in undiluted unit dose cartridges. This formulation has become more
predominant since the ban of CFC's in 1995 because of their ozone depleting
action. Achieving Sterility with EO Items
must be clean before sterilization. Materials to be sterilized, and packaging,
should be maintained in an environment with a relative humidity of at
least 50%. Items to be sterilized must be completely dry. All traces of
lubricants (exception is instrument lubricants which must be compatible
with EO) must be removed. Most packaging materials that are acceptable
for steam are also acceptable for EO.
Problems Associated With EO The benefits
of EO outweigh the risks. However, one must be aware of the health hazards
associated with EO residuals caused by improper or inadequate aeration,
which can cause death or irreversible tissue damage. Costs - ethylene
oxide gas is more expensive per cycle than steam and most of the alternative
sterilization methodologies due to the necessary compliance with OSHA
regulations. In addition, the extended time to properly and safely aerate
devices (a complete cycle can take 12-16 hours) makes turnover of devices
unrealistic.
This system
was cleared in 1993 as an alternative to ethylene oxide. It uses a small
amount of liquid hydrogen peroxide, which is energized with radio frequency
waves into gas plasma. Cycle time is 45-50 minutes depending on the load
configuration. The temperature is below 122oF. (50oC.).
The hydrogen peroxide is provided in multi-dose cassettes containing 10
single doses of liquid (nominal) 59% hydrogen peroxide. All devices processed in LTGP must be thoroughly cleaned and dried. Any moisture remaining in devices can result in an abort of the cycle. Compressed air can be used to blow out moisture out of lumens and other hidden places. As
with all devices, only those devices which meet the clearances for the
LTGP system and/or are cleared by the device manufacturer, should be processed
in this system.
The system only requires an electrical hook-up so installation is inexpensive. There are restrictions on lumen diameter and length. (NOTE: There is now a smaller unit that is compatible with most lumens). The phases of the basic LTGP include:
Packaging
materials - Instrument trays used in LTGP should be designed to optimize
diffusion of the hydrogen peroxide and not interfere with the Radio Frequency
(RF) energy or absorb hydrogen peroxide. Do not use linen, paper wraps,
peel packaging materials, or any cellulose-based material. Check with
tray manufacturers before purchase/use of containers. The following
packaging materials are compatible with LTGP; trays from the sterilizer
manufacturer; Tyvek (all plastic) pouches - no paper plastic pouches and
polypropylene based wraps. Ozone sterilization was cleared for use in the US in 2004. It uses oxygen that is subjected to an intense electrical field that separates oxygen molecules into atomic oxygen. It then combines with other oxygen molecules to form ozone. At the end of the cycle oxygen and water vapor safely vent into room, leaving no toxic residues.
Ozone Sterilization is easy to use. The sterilizer is 4.3 cubic ft in size. Ozone Sterilization cycle parameters (repeated twice):
Ozone Sterilization - requires a well ventilated room with 10 air exchanges/hour. Installation only requires an electrical outlet, oxygen source and demineralized water for humidification. The total cycle time is 4 hours.
LIQUID CHEMICALS Activated
Glutaraldehyde - is a liquid high-level disinfectant (however with prolonged
soak times sterilization can be achieved). Thorough pre-cleaning of devices
is required! It is mainly used for immersible items. It is critical that
the solution makes contact with all surfaces of the device to achieve
high-level disinfection (HLD). Follow the manufacturer’s instructions for use regarding temperature of the solution, etc. If a specific temperature is recommended, a thermometer is needed to verify the proper temperature has been achieved and maintained throughout the disinfection process. Today, the main use is for flexible and rigid scopes. It is non-corrosive to plastic, metal and lensed instruments. Glutaraldehyde has a shelf life (printed on the jug - determined by the manufacturer) and a use life. One formulation adds a buffer to make the solution last longer. Because the minimum effective concentration (MEC) of the solution can fall below the required level (even before the stated expiration date) it is now recommended that the efficacy of the solution be tested before each use. Test strips should be purchased from the solution manufacturer (more accurate). The minimum soak time is determined by the disinfectant manufacturer and is often affected by temperature. The use life
of the disinfectant can be adversely affected by soils, temperature and
in-use dilution (from not removing all rinse water after cleaning). Glutaraldehyde - The 2% alkaline solution is effective against all vegetative bacteria, viruses, TB and fungi. The solution can be toxic; therefore thorough rinsing is required to remove all residues. Three separate rinses should be performed and the water should not be reused. For cameras and rigid scopes, sterile water should be used. It is important to understand that the quality of the rinse water can re-contaminate the device. Most often, sterile water is used for the rinse. When using glutaraldehyde, proper ventilation is needed. The vapors can be a respiratory irritant. The liquid can be a skin irritant. The solution should only be used in a limited traffic area. A local exhaust hood may be needed to capture vapors during processing. The hood should be connected to a non-re-circulating exhaust system to the outside. There are also self contained systems. The hood should be monitored for efficacy. Special personal
protective equipment is needed when using glutaraldehyde. It is recommended
to wear eye shields, fluid resistant mask (for liquid not fumes), butyl
or nitrile rubber gloves (no vinyl or neoprene) and a polyethylene gown
with long sleeves. Store the solution covered (in a basin). A Spill Plan
is needed in the event the solution is spilled. For disposal, check the
local or state regulations. Usually, it is recommended that the glutaraldhyde
be diluted with copious amounts of running water. Dispose of containers
per label instructions. Ortho-phthaldehyde (0.55%) Cidex OPA Another high-level disinfectant is Cidex OPA. It is a non-glutaraldehyde product that is non-toxic and has a label claim for a 12-minute soak time for high-level disinfection at a minimum temperature of 20C (68F) for manual high level disinfection. No employee
monitoring is needed. It has a 14 day use life. Any unused portion can
remain in the original bottle for 70 days. Cidex OPA should be used in
a well-ventilated area. PPE is needed. You may have to double glove because
the product stains protein. This is seen as an advantage because if the
device is not properly the protein soils will stain grey/black. This product
is non-forgiving, you must use as directed.
Liquid Peracetic
Acid Processing System Monitoring the Cycles: ETO sterilization is monitored using internal and external chimerical indicators specifically designed to react to the parameters for ETO sterilization. The spore used to monitor ETO cycles is bacillus atrophaeus. Because of the critical parameters needed for sterilization, AAMI recommends biological monitoring of each cycle/load of ETO. LTGP is monitored using special internal and chemical indicators specifically designed to react to hydrogen peroxide. The chemicals used in the indicators are sensitive to fluorescent light; therefore they should be stored away from direct light (follow the manufacturer’s instructions for storage conditions). The spore used to biologically monitor LTGP is geobacillus stearothermophilus. AAMI recommends BI testing daily preferably with each load. The spore used to biologically monitor LTGP is geobacillus stearothermophilus. AAMI recommends BI testing daily preferably with each load. The Peracetic acid system requires a diagnostic test be performed daily before the system is placed into use. It is recommended to use a chemical indicator specific for peracetic acid in each cycle. Follow the manufacturer’s instructions for use and appropriate color change. The CI manufacturer’s instructions for use (including interpretation of the color change) and storage (including the expiration date) should be followed. At the end of the process if a CI suggests inadequate processing, the contents of the load should not be used. At this time there is no biological test for this system since it is a liquid chemical system. SUMMARY There are
a number of alternatives to steam sterilization available today. All have
advantages and disadvantages. It is important for an organization to extensively
review the alternatives and what would best meet the need of your organization.
Consider what types of surgical instrumentation you need to process. Two
very important considerations are space and cost. The space requirements
can be significant and you want to know what the process will cost in
both capital and operational expense. It is also important to involve
other members of the team in the purchase of this type of equipment. Often,
engineering or construction services will work closely with the vendor
to meet the requirements for the installation of a sterilization system.
This is important work to due prior to finalizing a purchase to reduce
implementation problems later on. References: TSO3 website for Ozone Sterilization. Association for the Advancement of Medical Instrumentation. AAMI/ANSI.ST-41 “Ethylene Oxide Sterilization in Health Care Facilities: Safety & Effectiveness” 2008. Association for the Advancement of Medical Instrumentation. Chemical Sterilization and High-Level Disinfection in Health Care Facilities. ANSI/AAMI ST58:2005. Arlington (VA): AAMI, 2005.
|