STERILE
PROCESSING UNIVERSITY
"Recalls"
Copyright © 2008 - Sterile Processing
University, LLC - All Rights Reserved.
This in-service may not be copied or used without permission of the author.
**This
site is best viewed at 1024x768 resolution on your monitor.
We
have all had to deal with recalls of one kind or another. It might have
been related to items sterilized in our own departments or with commercially
sterilized items. Regardless of the source, we must take immediate action
to account for all of the affected lot numbers.
Frequently,
within our own departments, the need to recall items is due to a positive
biological indicator (BI). The purpose of the BI is to challenge the sterilization
process. The test indicators validate that sterilizing conditions were
attained by killing the spores in the test pack. When a positive BI is
noted, the first step is to review all mechanical and chemical monitoring
records since the last negative BI. Only 10% of positive BI's are due
to equipment failure. Review the sterilization printout carefully to be
sure there were no inaccuracies. Before beginning the recall process,
look at the following carefully. These are all possible causes of a positive
BI:
- Was there
inadequate sterilization time?
- Was the
exposure time set correctly for the temperature?
- Review
how the sets were prepared, packaged, labeled, and loaded into the sterilizer.
- Was the
test pack improperly processed?
- Review
the actual processes performed by the staff to determine if the positive
was caused by the operator. 85% of most sterilization errors are caused
by the operator.
- Is it
possible that the load was never started and the contents were never
processed?
- Was the
correct BI used?
- Was the
control and the BI from the same lot?
- Was the
BI contaminated with microorganisms not originally in the BI?
- Was the
incubation faulty?
Recalls due
to positive biological indicators require all items from cycles sterilized
back to the last negative biological indicator results in the affected
sterilizer be retrieved. The first thing a staff member needs to do is
notify the appropriate supervisor or their designee of situations that
may warrant a recall such as a positive biological indicator, a failed
mechanical indicator, an unacceptable internal or external chemical indicator
or any situation that indicates that a load was not adequately processed.
The supervisor or designee would then make the decision to implement a
recall based on the facilities policies and procedures.
Recall procedures are established to expedite the retrieval of processed
items that have been determined to be suspect, immediately quarantine
the sterilizer and notify physicians or departments that may have received
the suspect items. Once a recall has been deemed necessary, notification
of physicians and departments should be the beginning of the recall process
and subsequent to needed documentation. The information used in the documentation
could include but is not limited to:
- Supervisor
or designee initiating and or authorizing the recall
- The individual
responsible for documenting the results of the recall
- Reason
for the recall
- Time
and date of the suspect cycle
- Description
of the load contents with reference given to the Lot and load control
numbers.
- The results
of the print out or graph depending on your sterilization unit.
All medical
devices processed in that specific sterilization unit since the last negative
BI are considered non-sterile. It is of utmost importance to accurately
assign lot numbers and record items to be sterilized within our own departments
because this is the record to be used to identify all items from affected
cycles. Once all items have been identified they are to be located and
retrieved. All must be completely reprocessed. A single positive BI does
not indicate a sterilizer malfunction if mechanical and chemical indicators
are acceptable and appear to be functioning properly. A second BI should
be immediately processed while holding the suspect items in quarantine.
If you would like to know if the BI was contaminated and human error caused
the positive, the positive BI can be sent to the lab for identification
of the specific microorganism. If it is different than the microorganism
in the BI you have confirmed that the BI has been contaminated and the
positive was caused by human error.
When there is evidence of a sterilization failure the Infection Control
Department should be notified to follow-up if any of the suspect medical
devices were used on patients and consult with the physicians that came
in contact with the suspect medical devices.
Notify your sterilizer service representative for assistance in determining
the sterilization failure. The combined results of mechanical, chemical
and biological monitoring should determine the success of any changes
made or become obvious if the sterilizer continues to malfunction. Once
the cause has been noted, arrange to have the unit serviced or repaired.
After the corrections have been made, a vacuum steam sterilizer must be
re-validated with three consecutive negative biological monitors in three
consecutive cycles, followed by three consecutive dynamic air removal
tests. Each type of cycle (gravity and pre-vacuum) must be tested and
all test results negative before the sterilizer is put back into use.
For table top sterilizers designed to be used with multiple types of modes
or cycles (e.g., 270F to 274F [132C to 135C], unwrapped instruments; 270F
to 274F [132C to 135C], wrapped instruments or peel pouches; 250F [121C],
wrapped packs; 250F [121C], liquids), then each sterilization mode or
cycle should be routinely tested with a BI PCD. The unit is not to be
used until the results of these tests are acceptable. Once the sterilizer
is determined to be functioning properly it can be put back into routine
use. It is important to keep records. This includes the time/date of the
cycle, sterilizer identification, packaging materials, location of the
BI, results of the test and the control, and the name of the person conducting
and reading the test.
It has been shown that the prevention of recall situations is frequent
biological monitoring and quarantine. Increased biological monitoring
can contribute to the reduction in healthcare acquired infections (HAI's),
cost reduction in decreased reprocessing and liability, and will promote
the confidence needed for good work flow in CS/SPD. Some facilities feel
it is clinically and operationally more effective to monitor every load
with a biological monitor than it is to deal with the costs associated
with recall, HAI's and litigation.
AAMI recommends that sterilizers be biologically monitored at least once
a week, preferably daily, when normal cycles are used, in each flash sterilization
load and in any load containing an implantable device. A positive BI is
a significant event. Different sterilization units are validated by different
microorganisms. A biological monitor specific to the type of sterilization
unit being monitored is necessary. Determine that the correct biological
indicator was used. Steam, low temperature gas plasma, and ozone BI's
contain Geobacillus stearothermophilus ( previously Bacillus
stearothermophilus). Bacillus atrophaeus (previously Bacillus
subtilis) is used to test ethylene oxide and dry heat sterilizers.
After processing, the BI's are incubated. If viable spores remain, they
will grow creating a positive biological. Biological indicators should
be handled and used according to the manufacturers instructions. Biological
monitors that test positive are to be discarded as medical waste.
AAMI Recommended Practices clearly state that our healthcare facilities
must have written policies and procedures for the recall of medical devices.
These policies and procedures are to be developed cooperatively with Infection
Control and Risk Management. They are to be documented and all records
maintained.
The recall outline recommended by AAMI (Comprehensive Guide to Steam
Sterilization and Sterility Assurance in Healthcare Facilities: ST-79,
2006) is as follows:
10.11.2
Recall Procedure
should
a) be written
b) outline the circumstances for issuing a recall order
c) designate the person(s) authorized to issue a recall order; and
d) designate the person(s) responsible for reporting on the execution
of a recall order
10.11.3
Recall Order should
e) include all items processed back to the last negative BI; be immediately
communicated to affected departments and followed by a written order;
f) identify by sterilization lot number the products to be recalled;
g) identify the persons or departments to whom the order is addressed;
h) require the recording, in terms of kind and quantity, of the products
obtained in the recall; and
i) specify the action to be taken by the persons receiving the order (e.g.,
destruction or return of the product)
10.11.4
Recall Report should
j) identify the circumstances that prompted the recall order
k) specify the corrective action(s) taken to prevent a recurrence
l) state, in terms of the total number of products intended to be recalled,
the percentage of products actually located in the recall; and
m) provide verification that the recalled items were reprocessed or destroyed,
as appropriate
REFERENCES:
AAMI ST:
79 2006, Comprehensive Guide to Steam sterilization and sterility assurance
in healthcare facilities. Section 10.11
The Basics
of Sterile Processing, second edition, 2007. Sterile Processing University,
LLC, Lebanon, NJ
|