|
STERILE PROCESSING UNIVERSITY "Table
Top Sterilizers" **This site is best viewed at 1024x768 resolution on your monitor. Table top sterilizers are usually found in facilities where there are a limited number of medical devices to be sterilized such as physician offices, dental offices, clinics and laboratories. A table top
sterilizer is a compact pressure vessel that has a chamber volume of not
more than 2 cubic feet that generates its own steam when distilled or
deionized water is added by the operator. Distilled or deionized water
is recommended to prevent buildup of minerals in the reservoir and on
processed devices and to ensure the purity of the steam generated for
sterilization. Table top sterilizers are typically "unplumbed"
i.e., not connected to a steam source and must be checked daily to ensure
that there is enough water in the reservoir for the number of loads to
be processed. The water is electrically heated and vaporized into the
chamber during the sterilization process by a valve or other means of
injection. After each cycle a condensation coil condenses the steam back
into water. Sterilization cycles are between 250 degrees F (121 degrees
C) and 270 degrees F (132 degrees C). Gauges (manual) or LED displays
(automatic) are used to display temperature and pressure. Mechanical timers
(manual) or programmable timers (automatic) are used to determine the
length of the sterilization cycle. Steam sterilizers
in general consist of: a sterilization chamber, a source of steam, a process
for evacuating air from the chamber and a method of infusing steam into
the chamber, a means of heating the water and controls for timing the
sterilization cycle. The sterilization chamber is the totally enclosed
area into which materials are placed for sterilization. This chamber is
equipped with an access door for closing and sealing (gasket) the chamber.
Prior to operating the sterilizer all personnel should be taught to properly
operate the sterilizer; the cycles it has available and what, if any,
maintenance is required. It is essential to read the manufacturers instructions
for operation and maintenance. The manufacturer's instruction manual should
be kept by the user as long as the sterilizer is in service. Before sterilization or high level disinfection, instruments should be thoroughly cleaned of all debris. Cleaning is an absolute pre-requisite to the sterilization of surgical and dental instruments. Thorough cleaning followed by rinsing is important to prevent spotting, staining and the accumulation of biofilms. To prevent drying or encrustation of debris, instruments should be kept moist with water or an enzymatic foam or gel if they cannot be cleaned immediately after a procedure. Enzymatic solutions facilitate effective instrument cleaning. Personal protective equipment must be worn when handling contaminated instruments. Packaging
cleaned, rinsed and dried instrumentation prior to sterilization protects
them from contamination after they are removed from the sterilizer and
during transport or in storage. The primary function of a packaging material
is to allow sterilization of the contents, to maintain the sterility of
the contents until the package is opened and to provide for the removal
of the contents without contaminating it. It is important to follow the
manufactures instructions for cleaning and sterilization of medical devices
to ensure the effectiveness of the selected sterilization process. All
instruments should be held open and in the unlocked position. Instruments
that can be disassembled are to be processed as separate pieces. Devices
with stylets or obturators should be disassembled before processing them
in a table top steam sterilizer. Most, but not all, tabletop steam sterilizer
manufacturers have not validated sterilization of either double peel pouches
or peel pouches within wrapped goods. If the sterilizer you are using
has not been validated for these configurations, the practice should be
considered unacceptable because there is no assurance that the sterilizer
is capable of processing the items. When removing items from the sterilizer they should be visibly dry. Packages should not be touched while they are hot. Condensate moisture can wick contaminants into the package. Sterile items should never be handled before they are cool. Once the items are removed from the sterilizer they should be allowed to cool in the loading tray untouched. Place the tray in a low traffic area, away from air vents to decrease exposure to the environment. After the sterile items are cool they should be stored in a clean, dry, easily accessible area on or in designated shelving, counters or containers. They should never be stored under sinks, near water or sewer pipes, or in any area where they can become wet. Qualification
testing should be done before the sterilizer is put into regular use to
establish a baseline for the sterilizer's performance. The same testing
should be done if there has been a sterilizer malfunction or a sterilization
process failure. For the gravity displacement steam sterilizer validation
is performed by performing three biological test packs processed back
to back in full loads. Test packs should be assembled in a way that simulates
the items routinely sterilized. If the tabletop is a pre-vacuum steam
sterilizer the three biological tests should be followed by three dynamic
air removal tests, processed back to back. The BIs in the pre-vacuum cycles
should also be in full loads and packaged to simulate the items routinely
sterilized. If large wrapped trays will be processed it may be possible
to use commercially prepared test packs. Check with the BI manufacturer
to determine if their packs are compatible with the table top sterilizer
cycles you will be using. If multiple cycles are used (e.g. wrapped,
unwrapped,) all must be tested three times. All three of the tests
must be negative. Only after the results are known can the sterilizer
can be put in to use. AAMI recommends
that routine biological testing be performed at least weekly but preferably
daily. Each load containing implantable devices should be monitored and,
when possible, the implantable device should be quarantined until the
monitoring results have been obtained. If a sterilizer is to be used for
multiple types of cycles each cycle should be monitored at least weekly.
The more often the sterilization process is monitored the more likely
you are to detect a sterilization process failure. These simple precautions
reduce the risk of a patient infection and the financial impact of a recall,
as well as cost of reprocessing. Proper functioning of steam sterilization
cycles should be verified by the routine use of biological indicators.
Biological indicators consist of about one million highly resistant bacterial
spores of Geobacillus stearothermophilus. Several types of biological
indicators are available. It is imperative that the BI used is the one
designated for the cycle you are testing. The sterilizer manufacturer
determines the placement of the biological. Generally, it is placed on
the bottom shelf toward the front of the sterilizer chamber. Processed
items should be held in quarantine until the results of the BI is known.
A control biological indicator is necessary to validate the pre-sterilization
viability of the test spores; the lot control number of the BI vial must
match the lot control number of the control vial. The BI manufacturer's
instructions should be followed for storing, using, incubating and reading
the BI. All biological indicator results, including results from controls,
should be interpreted by a qualified individual and the results noted
in the sterilization records. Positive biological indicators should be
disposed of according to manufacturer's instructions. Mechanical
monitoring is best described as the charts, gauges and printouts unique
to each sterilization unit. At the end of each cycle the user should check
and verify that the time, temperature and pressure were correct. These
records should be initialed by the user and any discrepancies should be
reported. If tabletop sterilizers do not have some type of recording device
one should be installed. If the sterilizer cannot be equipped with a printout,
a Class V chemical integrator should be used with each load to provide
additional documentation that cycle parameters were met. The information
supplied here does not address all that one needs to know about table
top sterilizers but it does provide an overview of what they are, how
they are used and how they are monitored. For more information regarding
Table Top Sterilizers refer to AAMI ST-79 (Comprehensive Guide to Steam
Sterilization and Sterility Assurance (2006) and the Basics of Sterile
Processing, 2nd Ed., 2007.
|